MetriScan uses radiographic absorptiometry (RA) to estimate relative phalangeal bone density of the three middle fingers. With RA a high resolution radiographic image of a patient's phalanges is taken. A computerized analysis is made comparing the intensity of the image with a reference wedge embedded under the hand plate.
Validation
This method was validated in one of the largest epidemiological studies in the bone density field*. Phalangeal bone density testing was performed on 3481 women, who were then followed for sixteen years. The study concluded that phalageal bone densitometry is a significant predictor of future hip fractures.
Features
• 0.3 second scan time
• Radiation dose is less than 5% of a dental x-ray
• Precision error = 1.1%
• Automatic calibration check
• Small footprint: 16" x 16" x 16"
Specifications
| Weight |
42 lbs |
| Dimensions |
16" x 16" x 16" |
| Precision Error |
1.1% CV (in vivo) |
| Patient x-ray dose |
0.001 mrem (effective
dose equivalent) |
| Mains voltage |
100 - 240 vac |
| Mains frequency |
50 - 60 Hz |
| Printer |
HP, as specified by Alara |
* Mussolino, M., et al, Archives of Internal Medicine, 1997; 157:433-438
Class I Laser
Product |