MetriScan uses radiographic absorptiometry (RA) to estimate relative phalangeal bone density of the three middle fingers. With RA a high resolution radiographic image of a patient's phalanges is taken. A computerized analysis is made comparing the intensity of the image with a reference wedge embedded under the hand plate.

Validation
This method was validated in one of the largest epidemiological studies in the bone density field*. Phalangeal bone density testing was performed on 3481 women, who were then followed for sixteen years. The study concluded that phalageal bone densitometry is a significant predictor of future hip fractures. 

Features
• 0.3 second scan time
• Radiation dose is less than 5% of a dental x-ray
• Precision error = 1.1%
• Automatic calibration check
• Small footprint:  16" x 16" x 16"

Specifications

 Weight  42 lbs
 Dimensions  16" x 16" x 16"
 Precision Error  1.1% CV (in vivo)
 Patient x-ray dose  0.001 mrem (effective dose equivalent)
 Mains voltage  100 - 240 vac
 Mains frequency               50 - 60 Hz
 Printer  HP, as specified by Alara

* Mussolino, M., et al, Archives of Internal Medicine, 1997; 157:433-438

Class I Laser Product